G70G80 Patient Monitor

G70/G80 Modular Patient Monitor

Philips Goldway G70 & G80 Patient Monitors


Product Description

  • Plug-and-play modular design provides patient care flexibility
  • 15″/17″ color TFT display
  • Maximum 10/12 waveforms display
  • 7-lead ECG or 12-lead ECG simultaneous display
  • ST and arrhythmia analysis
  • Hemodynamic, Oxygenation, Ventilation, Drug Dose Calculations
  • OxyCRG dynamic trend display
  • Bed-to-bed view
  • Applications from neonate to adult
  • LAN networking capability to UT4800 central monitoring system

Additional Information

Multi Parameter Module

dual-slot

ECG (Philips ECG Technology):

Lead type: 3-lead or 5-lead selectable, optional
12-lead
Lead selection:
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1,
V2, V3, V4, V5, V6

HR Measurement Range:

Adult/Pediatric: 15 bpm – 300 bpm
Neonate: 15 bpm – 350 bpm
Accuracy: ±1 bpm or ±1 %, whichever is greater

ST measurement range: -2.5 mV – +2.5 mV
Accuracy: ±0.05 mV or ±15 %, whichever is greater

Arrhythmia: 21 calls

Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50mm/s

Gain selection: x 1/4, x 1/2, x 1, x 2, x 4, Auto

Pacemaker detection: Yes

Protection: Against electrosurgical interference and defibrillation.

ECG analog output

Non Invasive Blood Pressure (NIBP)

Method: Automatic oscillometric
Parameters: SYS/DIA/MAP/Pulse Rate
Operating mode: Manual/Auto/STAT

Auto measure intervals:
1-10, 15, 30, 60, 90, 120, 180, 240, 480 minutes

Units: mmHg/kPa

Cuff pressure range:
Adult/Pediatric: 0 – 300 mmHg (0 kPa – 40.0 kPa)
Neonate: 0 – 150 mmHg (0 kPa – 20.0 kPa)
Measurement and alarm range:

SYS: Adult/Pediatric: 30 – 255 mmHg (4.0 kPa – 33.9 kPa)
Neonate: 30 – 135 mmHg (4.0 kPa – 18.0 kPa)
DIA: Adult/Pediatric: 10 – 220 mmHg (1.3 kPa – 29.3 kPa)
Neonate: 10 – 110 mmHg (1.3 kPa – 14.7 kPa)
Accuracy: Max. Mean Error: ±5 mmHg (±0.7 kPa)

Max. Std. Deviation: 8 mmHg (1.1 kPa)
PR measurement range: 40 bpm – 240 bmp
Accuracy: ± 2 bpm or ±2 %, whichever is greater
Over-pressure protection: Yes

SpO2 (Philips SpO2 Technology)

Measurement range: 0 – 100 %
Accuracy: 70 – 100 %, depends on sensor
0 – 69 %, unspecified

Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50mm/s

PR measurement range: 30 – 300 bpm
Accuracy: ±1 bpm or ±2 %, whichever is greater

Respiration (RESP)

Measurement method: Thoracic impedance
Measurement range: Adult: 0 – 120 rpm
Pediatric: 0 – 150 rpm
Neonate: 0 – 170 rpm

Accuracy: ±1 rpm in the range of 0 – 70 rpm, ±2 rpm in the range of 71 – 120 rpm, ±3 rpm in the range of 121 – 170 rpm

Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s

Gain selection: x 1/2, x 1, x 2

Lead selection: RA – LA, RA – LL, LA – RL, LL – RL

Apnea alarm: Yes, 10 – 60 s

Temperature (TEMP, 2 Channel)

Measurement range: 0°C – 50°C (32°F – 122°F)
Resolution: 0.1°C (0.2°F)
Accuracy: ±0.1°C (±0.2°F)
Units: °C/°F

IBP (2 Channel, Optional)

Parameters: SYS/DIA/MAP/Pulse Rate
Measurement range: -50 – 400 mmHg (-6.6 kPa – 53.2 kPa)
Accuracy: ±4 mmHg (±0.53 kPa) or ±4 % of
reading, whichever is greater
Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s,
50 mm/s
Resolution: 1 mmHg (0.13 kPa)
Units: mmHg/kPa
Transducer sensitivity: 5 uV/V/mmHg

EtCO2 Module (single Slot) (optional)

Method: Microstream, infrared absorption
Parameters: RR, etCO2, inCO2
Measurement range: 0 – 99 mmHg (0 – 13.2 kPa)
Accuracy:
0 – 38 mmHg (0 kPa – 5.06 kPa): ±2 mmHg (±0.3 kPa)
39 – 99 mmHg (5.19 kPa – 13.2 kPa): ±5 % of reading + 0.08 %
for every 1 mmHg above 38 mmHg (5.2 kPa – 13.2 kPa: ±5 % of
reading + 0.08 % for every 0.13 kPa above 5.1 kPa)
Resolution: 1 mmHg (0.13 kPa)
Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
Respiration rate: 0 – 150 rpm
Accuracy: ± 1 rpm in the range of 0 – 70 rpm
± 2 rpm in the range of 71-120rpm
± 3 rpm in the range of 121 – 150 rpm
Units: mmHg/kPa

ECG Analog Output (optional)

Parameters: SYS/DIA/MAP/Pulse Rate
Measurement range: -50 – 400 mmHg (-6.6 kPa – 53.2 kPa)
Accuracy: ±4 mmHg (±0.53 kPa) or ±4 % of
reading, whichever is greater
Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s,
50 mm/s
Resolution: 1 mmHg (0.13 kPa)
Units: mmHg/kPa
Transducer sensitivity: 5 uV/V/mmHg
Cardiac Output Module (single-slot)
Measurement method: Right heart thermodilution
Measurement range: 0.1 l/min – 20.0 l/min
Accuracy: ±0.2 l/min or ±10 %, whichever is
greater
Blood temperature range: 27°C – 43°C (80.6°F – 109.4°F)
Accuracy: ±0.1°C
Injection Temperature: 0°C – 27°C (32°F – 80.6°F)
Units: °C/°F

Multi Gas Module (triple Slot) (optional)

(Manual Agent ID or Auto Agent ID with O2)
Waveform speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50
mm/s
O2 measurement range: 0 Vol.% – 100 Vol.%
Accuracy: 0 – 40: ± 1 Vol.%
40 – 60: ± 2 Vol.%
60 – 80: ± 3 Vol.%
80 – 100: ± 4 Vol.%
N2O measurement range: 0 Vol.% – 100 Vol.%
Accuracy: 0 – 20: ± 2 Vol.%
20 – 100: ± 3 Vol.%
CO2 measurement range: 0 % – 10 % (0 mmHg – 100 mmHg)
Accuracy: 0 – 1: ± 0.1 %
1 – 5: ± 0.2 %
5 – 7: ± 0.3 %
7 – 10: ± 0.5 %
> 10: unspecified
Enflurane / Isoflurane/ Halothane measurement range:
0 Vol.% – 5 Vol.%
Accuracy: 0 – 1: ± 0.15 %
1 – 5: ± 0.2 %
> 10: unspecified
Sevoflurane measurement range: 0 Vol.% – 8 Vol.%
Accuracy: 0 – 1: ± 0.15 %
1 – 5: ± 0.2 %
5 – 8: ± 0.4 %
> 8: unspecified
Desflurane measurement range: 0 Vol.% – 18 Vol.%
Accuracy: 0 – 1: ± 0.15 %
1 – 5: ± 0.2 %
5 – 10: ± 0.4 %
10 – 15: ± 0.6 %
15 – 18: ± 1%
> 18: unspecified

Flexible Module Server (optional)

8 module slots
1.1m connection cable to monitor

Recorder (optional)

Type: Built-in thermal recorder, 3 channels
Print mode: Real-time recording, timed recording,
alarm-activated recording
Print speed: Auto, 12.5 mm/s, 25 mm/s, 50 mm/s

Display

Type: Color TFT LCD
Size: 15″/ 17″
Resolution: 1024 x 768 pixels (G70)
1280 x 1024 pixels (G80)
Channels: Max.10/12 channels

Power Supply

Power Supply
AC power input: 100 V – 240 VAC, 50/60 Hz
Battery (optional): Li-Ion battery, ≥2 hours of continuous
monitoring (with ECG, SpO2
measurement and NIBP measured
every 15 minutes)

Weight & Dimensions:

Main unit net weight: Approx. 9kg/10kg (with Multiparameter
Module, without recorder
and battery)
Main unit dimensions: 400 mm (W) x 192 mm (D) x 366 mm (H)
FMS net weight: Approx. 2 kg (without modules)
FMS dimensions: 437 mm (W) x 138 mm (D) x 134 mm (H)

Safety Approval & Quality System

Safety standard: IEC 60601-1
Quality system: ISO 9001 & ISO 13485 certified
CE marking according to the Council Directive 93/42/EEC

Data Review And Storage:

1600 sets trend table
1200-hour trend data of all parameters (with Data Manager
function)
120-minute full disclosure ECG 1 waveform review (with Data
Manager function)
200 sets alarm events
200 sets arrhythmia events

External Interface (standard)

Digital video output, FMS interface, Nurse Call, LAN, USB